Flow Rates
|
0.01 to 24.00 mL/min
|
Pressure |
0 to 10000 psi |
Pressure Accuracy
|
± 2% of full scale pressure
|
Pressure Zero Offset
|
0 psi +10 psi
|
Flow Accuracy (Constant Flow Pump)
|
± 2% using 100% Methanol @ 1000 psi
± 5% using CO2 (Based on gas volume measurements and back calculation to liquid)
|
Flow Precision (Constant Flow Pump)
|
0.5% RSD using 100% Methanol @ 1000 psi
|
Pressure Control Accuracy (Constant Pressure Pump)
|
<10% or 100 psi whichever is larger
|
Dimensions
|
5.75” high x 11.125” wide x 21.6” deep (15 x 28 x 54.9 cm)
|
Weight
|
24.0 lbs (10.8 kg)
|
Power
|
110-120 VAC, 50-60 Hz; or
220-240 VAC, 50-60 Hz
|
Back Panel Inputs
|
Run/stop inputs (5 volt TTL type)
0 to 10 volt flowrate control input
0 to 10 KHz flowrate control input
USB and RS-232 serial communication ports for complete control and status monitoring
|
Back Panel Output
|
Normally open and normally closed relay contacts (SPDT, form C, 0.25 amp max, 50 volt max) indicate when a pressure fault or motor stall fault occurs.
|
Pulsation (Constant Flow Pump)
|
± 5% @ 1.0 mL/min using 100% Methanol @ 1000 psi
|